Lab console · internal

Results entry — chromatography

Import the detector export for a run, integrate it, and compare the measured trace against the compound's reference profile (the expected retention behavior on the house method). The output is a batch record whose chromatogram is the instrument's own data.

Data integrity rule. The certificate figure is rendered only from an imported instrument export, and every batch record carries the SHA-256 of that source file. Reference profiles are expected-value overlays derived from method calibration — they are drawn dashed, labeled, and never stored or displayed as a measurement.

1 · Sample

Identify the lot under test. The compound selection loads the matching reference profile (expected RT, acceptance window, named impurities) from peptides.json.

Reserved in the lab database — duplicates are impossible.

2 · Instrument export

Drop the chromatogram export from the data system — Agilent ChemStation/OpenLab CSV, Shimadzu LabSolutions ASCII, EZChrom-style single-column ASCII, or any text export (.exp included) with a time/signal table or a sampling-rate header. The file is parsed locally; nothing is uploaded. Binary native files can't be read — use the data system's text/ASCII export.

The example is synthetic and watermarked — records built from it cannot be exported.
Screenshots are curve-digitized; vendor / instrument COA PDFs are read and regenerated as our certificate. Review everything against the original.

3 · Trace & integration

The solid trace is the imported run. The dashed overlay is the reference profile for the selected compound, scaled to the main peak for visual comparison; the shaded band is the RT acceptance window. Adjust integration below — excluded peaks are kept in the record, marked excluded.

Measured trace (instrument export) Reference profile (expected — not a measurement) RT acceptance window

4 · Checks

Computed from the integrated trace against the reference profile and house specs. A FLAG does not block the record — it is recorded as a flag for analyst review.

5 · Batch record

The record bundles the raw trace, the source-file hash, integration results, checks, and sample metadata. It is the artifact the certificate and the public Verify page are generated from.

Import an instrument export and fill in lot, report ID and analyst to enable.